Description
SPECIFICATIONS :
GN-10-II is a dual stainless steel glove box designed for researchers in material science, chemistry, semiconductor, and related topics. The purification cylinder is made of BASF antioxidant materials from Germany and UOP water adsorbent materials from the US in order to achieve a low water and oxygen concentration of lower than 1ppm, such that experiments can be conducted in an ultra clean and highly purified gas environment. It is integrated with vapor pressure control system, purity control system, automatic gas purity regeneration control system, auto alarm, and system data log recording. GN-10-II is widely used in preparing nanomaterials, Li-ion battery materials, catalysts, and metallorganics.
SPECIFICATIONS
1. Working Voltage:110VAC or 220VAC, 50/60Hz, Single Phase
2. Max. Power Consumption: 3KW
3. Glove Box Chamber :
- Openable left-front window panel (tempered glass, 8 mm thick)
- Dimensions: 2400mm(L) x 800mm(W) x 930mm(H)
- Case Material: Stainless Steel 304
- 12mm thick double layer laminated safety glass with corrosion protection film for operating in a safe and easy manner.
- <0.05 Vol% / hr (Depending on oxygen concentration inside the glove box)
- Built-in fluorescent lamp
- Two slidable shelves, 300mm L x 250mm W x 300mm H
4. Airlock Chamber :
- Antechamber Dimensions: 385mm(Dia.) x 588mm(L)
- Small Chamber Dimensions (for tools): 150mm(Dia.) x 300mm(L)
- Max. vacuum level: 100 Pa
5. Attainable purity level
- Water concentration: <1 ppm (20°C, 1 atm)
- Oxygen concentration: <1 ppm (20°C, 1 atm)
6. Working Gas
- Operating gas: Inert gas such as N2, Ar, He
- Regenerating gas: Mixture of H2 (5%) and Operating gas
7. Gas Purification System
- Auto regeneration process
- Auto purging function
- Purification Cylinder: made of BASF antioxidant materials and UOP water adsorbent materials
- Moisture removal capacity: 1.5KG
- Oxygen removal capacity: 180L
- Stainless steel pipe connection
- Blower: Flow Rate: 90m³/h
8. Filter System
- One Organic Solvent Filter System filled with replaceable activated carbon (5.2kg)
- Two pieces of HEPA filters can eliminate particle size > 0.3μm
9. Control system
· Siemens PLC control system with 6″ touchscreen display
· Water concentration: 0 – 1000 ppm @0.1 ppm
· Oxygen concentration: 0 – 1000 ppm @0.1 ppm
· Auto pressure control -25 to 25 mbar @0.1mbar
10. Feedthroughs
- Power feedthrough is included.
- power extension board with 5 outlets, 15A max.
- Can be 110VAC or 220VAC depending on the input source
11. Vacuum Pump
- 240L/m Heavy Duty Double Stage Rotary Vane Pump for faster gas exhaust
- Vacuum rate: 7.8 cfm
- Lowest pressure: 0.4pa
12. Gloves
- A pair of NORTH Butyl glove is included
- protection against toxic materials
- 8″ Dia.x 32″ Long
13. Product Dimensions:3115mm(L) x 830mm(W) x 1830mm(H) (123″ x 33″ x 72″)
14. Warranty:
- One Year limited warranty with lifetime support
- Rusting and damage due to improper storage condition or maintenance are not covered by warranty
- Gloves are consumable items and are not covered under warranty
15. Application Notes & Warnings:
- The interconnections between the glove box chamber and the gas purification system must be unimpeded during the purification cycles.
- The use of corrosive gasses is prohibited as they will damage the water and oxygen sensors.
- Regularly perform regeneration to maintain the optimum purification efficiency.
- The O2 removing rate is largely dependent on the type of purging gas used. For faster chamber purging process, Nitrogen is preferred over Argon due to its lighter mass.
- Corrosive liquid (such as LiPF6 electrolyte) must be remain sealed in a container inside the glove box. Otherwise, liquid vapor may condense and corrode the steel chamber and purification pipeline
- The golvebox may connect with tube furnace upon request